Lamotte Tracer Total Chlorine Meter User Manual

Browse online or download User Manual for Equipment Lamotte Tracer Total Chlorine Meter. LaMotte Tracer Total Chlorine Meter User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 16
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
TRACER
TOTAL CHLORINE
TCl
CODE 1740
POCKETESTER
ä
Page view 0
1 2 3 4 5 6 ... 15 16

Summary of Contents

Page 1 - POCKETESTER

TRACERTOTAL CHLORINETClCODE 1740POCKETESTERä

Page 2

Total Chlorine Tracer Offset Adjustment Procedure1. Obtain a sample with a known concentration of total chlorine.2. Follow the Chlorine Measurement pr

Page 3 - ...10

3. Pipet exactly 1.00 mL of the 100 ppm solution into a third 100 mLvolumetric flask. Dilute to the 100 mL line with deionized water. Capand mix. This

Page 4

MAINTENANCEStorageWhen not in use, the TRACER should be stored dry with the electrodecap in place. Cover the connector with the clear cover whenever t

Page 5

Replacing the ElectrodeThe TRACER is shipped with an electrode attached. If the electrodeneeds to be replaced or changed, follow these steps.1. Press

Page 6

TROUBLESHOOTINGProblem Check ActionPower on but nodisplayBatteries Insert batteriesVerify correct polarityReplaceUnstable reading Electrode Immerse pr

Page 7

ORP TRACER Electrode,±999 mV Code 1734The ORP TRACER Electrode (code1734) requiresan initial soaking in a pH 4.0 buffer solution. OrderpH 4.0 Mini Buf

Page 8

LaMOTTE COMPANYHelping People Solve Analytical Challenges®PO Box 329 • Chestertown • Maryland • 21620 • USA800-344-3100 • 410-778-3100 (Outside U.S.A.

Page 9

WARNING! This set contains chemicalsthat may be harmful if misused. Readcautions on individual containerscarefully. Not to be used by childrenexcept u

Page 10

TRACERTOTAL CHLORINE POCKETESTERTMCODE 1740TABLE OF CONTENTSIntroduction ... 4Specifications ...

Page 11

TABLE OF CONTENTS (con't)Troubleshooting ... 14Expanding Your TRACER ...15Warranty ...

Page 12

SPECIFICATIONSDisplay Multifunction display with bar graphOperating conditions32 to 122°F(0to50°C) and < 80% RHChlorine range 0.01 to 10.00 ppm (To

Page 13

METER DESCRIPTIONFront Panel Description1. Battery compartment2. Display3. MODE/HOLD button-store data4. CAL/RECALL button-change temperatureunits and

Page 14

BASIC OPERATIONPowering the TRACERIf the batteries are weak, the BAT indicator will appear on the display.Press the ON/OFF key to turn the TRACER on o

Page 15

TESTINGConditioning the ElectrodeThe TRACER electrode comes in a factory cleaned condition. If themeter is new or has been in storage for an extended

Page 16 - LaMOTTE COMPANY

Recalling Stored ReadingsNOTE: The HOLD symbol should not be displayed. If it is, exit theHOLD function by pressing the MODE button.1. Press the CAL/R