Lamotte 2020i Turbidimeter User Manual

Browse online or download User Manual for Equipment Lamotte 2020i Turbidimeter. LaMotte 2020i Turbidimeter User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 44
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
2020e • Code 1979-EPA
2020i • Code 1979-ISO
Version 1.5 • Code 1979-MN • 12-07
e
/i
Page view 0
1 2 3 4 5 6 ... 43 44

Summary of Contents

Page 1 - 2020i • Code 1979-ISO

2020e • Code 1979-EPA2020i • Code 1979-ISOVersion 1.5 • Code 1979-MN • 12-07e/i

Page 2

SELECTING A TURBIDITY CALIBRATION CURVE1. Press ON to turn the meter on.1.32. Scroll down and then press *IOK toselect Options.Main MenuMeasureData Lo

Page 3

The Japan Standard calibration mode should be used only with JapanesePolystyrene Standards (0–100 NTU).Note: StablCal®standards below 50 NTU should no

Page 4 - 4 CONTENTS

4. Scroll down and then press *IOK toselect an averaging option.Available options are:Disabled, 2 Measurements,5 Measurements,10 Measurements.Note: Th

Page 5 - CONTENTS 5

4. Press *IOK to select Set Date orscroll down and then press *IOK toselect Set Time or Date Format.Date/Time*Set DateSet TimeDate Format16:02:19 01/0

Page 6 - TURBIDITY

SELECTING A LANGUAGEThe default setting is English. To change the setting:1. Press ON to turn the meter on.1.32. Scroll down and then press *IOKto sel

Page 7 - SAMPLE DILUTION TECHNIQUES

SETTING AUTO SHUTDOWNThe power saving Auto Shutdown feature will turn the meter off when a button has notbeen pushed for a set amount of time. The def

Page 8 - FACTORY DEFAULT SETTINGS

DATA LOGGINGThe default setting for the data logger is start (on). The meter will log the last 4004data points.1. Press ON to turn the meter on.1.32.

Page 9

Data LoggingView*StartErase16:02:19 01/04/05Or scroll down and press *IOK toselect Erase to empty all loggeddata points from the memory.Data LoggingVi

Page 10

TUBE POSITIONING RINGThe Tube Positioning RingTo put a ring on a tube, remove the cap from the tube.The tube positioning ring has two tapered notches

Page 11 - AVERAGING

ANALYSIS1. Press *ION to turn the meter on.1.32. Press *IOK to select Measure. Main Menu* MeasureData LoggingOptions16:02:19 01/04/053. Rinse a clean

Page 13 - 01/04/05

6. Open the meter lid. Insert the tubeinto the chamber. Align the indexnotch on the positioning ring withthe index arrow on the meter. Closethe lid.7.

Page 14 - SELECTING A LANGUAGE

10. Cap the tube. Wipe the tubethoroughly again with a lint-freecloth.11. Open the meter lid. Insert the tubeinto the chamber. Align the indexnotch on

Page 15 - SETTING AUTO SHUTDOWN

Dilution ProceduresIf a sample is encountered that is more than 4000 NTU, a careful dilution with 0 NTUor very low turbidity water will bring the samp

Page 16 - DATA LOGGING

2. Pour approximately 50 mL of deionized water into thebarrel of the syringe. Insert the plunger. Exert pressureon the plunger to slowly force the wat

Page 17

Turbidity StandardsOnly use AMCO or formazin standards with the 2020. StablCalâ standards below 50NTU should not be used to calibrate the 2020. The di

Page 18 - ANALYSIS & CALIBRATION

4. Fill the tube to the fill line with theblank. Pour the blank down theinside of the tube to avoid creatingbubbles. Cap the tube.5. Dry the tube with

Page 19 - ANALYSIS & CALIBRATION 19

8. Rinse a clean tube (0290), or thesame tube, three times with thestandard. Avoid spilling standard onthe outside of the tube.IMPORTANT: Whilethetube

Page 20 - 20 ANALYSIS & CALIBRATION

12. Press *IOK to select Scan Sample. Turbidity (F)* Scan Sample16:02:19 01/04/0513. Observe the result. Turbidity (F)0.54 NTU* Scan Sample q16:02:19

Page 21 - 0.54 NTU

Calibrate00.50q , Q , p16:02:19 01/04/0516. When the value on the displaymatches the concentration of theturbidity standard, press the *IOKto select S

Page 22 - 22 ANALYSIS & CALIBRATION

Testing Tips1. Samples should be collected in a clean glass or polyethylene container.2. Samples should be analyzed as soon as possible after collecti

Page 23 - ANALYSIS & CALIBRATION 23

WARNING! This set contains chemicalsthat may be harmful if misused. Readcautions on individual containerscarefully. Not to be used by childrenexcept u

Page 24 - 24 ANALYSIS & CALIBRATION

TROUBLESHOOTING GUIDEError MessagesErr1 Very Low battery. Replace battery or switch to AC power. Press backarrow () to back out. Scan sample again. R

Page 25 - ANALYSIS & CALIBRATION 25

Meter freezes Lid was opened whenreading was being taken.Close lid. Read again.Unplug adapter. Plugadapter in to reset.Stray LightThe accuracy of read

Page 26 - 26 ANALYSIS & CALIBRATION

32 GENERAL OPERATING INFORMATIONTopViewBottomViewSerialNumberBatteryCompartmentONOFFOKLMott2020

Page 27 - ANALYSIS & CALIBRATION 27

General Operating InformationThe operation of the 2020 is controlled by the menu driven software and userinterface. A menu is a list of choices. This

Page 28 - 28 ANALYSIS & CALIBRATION

The Display & MenusThe display allows menu selections to be viewed and selected. These selections instructthe 2020 to perform specific tasks. The

Page 29 - ANALYSIS & CALIBRATION 29

Header Menu TitleMessage or ResultWindowMessageor ResultMenu Selection Window * First Choice~Footer 16:02:19 01/04/05Second ChoiceanotherandsoonAs de

Page 30 - Troubleshooting

TubesThe 2020 uses one type of tube (Code 0290) for the turbidity test. There is no needfor a special turbidity tube.The handling of the tubes is of u

Page 31 - Overview

PC LINKThe 2020 may be interfaced with any Windows-based computer by using the LaMotteSMARTLink 2 Program and Interface Cable (Code 1912-3 with 3.5 in

Page 32 - Compartment

MAINTENANCECleaningClean meter with a damp, lint-free cloth.DO NOT ALLOW WATER TO ENTER THE METER CHAMBER OR ANY OTHER PARTSOF THE METER.Clean meter c

Page 33 - The Keypad

GENERAL INFORMATIONPackaging and DeliveryExperienced packaging personnel at LaMotte Company assure adequate protectionagainst normal hazards encounter

Page 34 - The Display & Menus

CONTENTSIntroduction...6Turbidity ...6WhatisTurbidity?...6HowisTurbi

Page 35 - > Over range indicator

Specifications - 2020e and 2020iInstrument Type: NephelometerStandard: EPA 180.1, 2020eISO7027, 2020iUnits of Measure& Range:NTU (Nephelometric Tu

Page 36 - 36 TUBES

Statistical and Technical Definitions Related to Product SpecificationsMethod Detection Limit (MDL): “The method detection limit (MDL) is defined as t

Page 37

Contents and Accessories2020e KitEPA VersionCode 1979-EPA2020i KitISO VersionCode 1979-ISOCONTENTS Code Code0 NTU Standard, 60 mL 1480 14801 NTU Stand

Page 39 - Limits of Liability

LaMotte Company • PO Box 329 • 802 Washington AvenueChestertown • Maryland • 21620 • USAUSA 800-344-3100 • 410-778-3100 • fax 410-778-6394 • www.lamot

Page 40 - 40 SPECIFICATIONS

Computer Connection ...37BatteryOperation ...37ReplacingtheBattery...37Mainten

Page 41

INTRODUCTIONTURBIDITYWhat is Turbidity?Turbidity is an aggregate property of the solution, water in most cases. Turbidity is notspecific to the types

Page 42 - Warranty

compared with the intensity passing through a turbidity-free sample at 180° from thelight source. This method is good for highly turbid samples. The m

Page 43

All dilutions are based on a final volume of 10 mL so several dilutions will requiresmall volumes of the water sample. Graduated pipets should be used

Page 44 - Code 1979-MN

2. Scroll down and then press *IOKto select Options.Main MenuMeasureData Logging* Options16:02:19 01/04/053. Scroll down and then press *IOKto select