LaMOTTE COMPANYHelping People Solve Analytical Challenges®PO Box 329 • Chestertown • Maryland • 21620 • USA800-344-3100 • 410-778-3100 (Outside U.S.A.
AN INTRODUCTION TO COLORIMETRIC ANALYSISMost test substances in water are colorless and undetectable to the human eye.In order to test for their prese
THE MODEL DC1100 COLORIMETERThe DC1100 colorimeter has been specially calibrated to read the test resultdirectly in parts per million (ppm) of a desir
GENERAL OPERATING INSTRUCTIONS1. Fill the water sample bottle (0688) with the sample water.2. Fill colorimeter tube (0967) to 10 mL line with water fr
5. For best results, use the same tube that was used to adjust the blank for allmeasurements. If the test you are performing requires a wait of more t
REAGENT BLANKFor greater accuracy, the test procedure should be run on a 10 mL sample ofdeionized water, known as the reagent blank. This procedure wi
TROUBLE SHOOTING GUIDEPROBLEM CHECK ACTIONMeter won’tcome onBattery? Replace, see pg. 6 of manual.^ AC Adapter? See pg. 6 of manual.^ AC Wall Outlet?
TROUBLE SHOOTING GUIDE (continued)PROBLEM CHECK ACTIONSuspectCalibrationVerify standard withalternate test method?Run alternate test on standard.^ Ver