HEXAVALENT CHROMIUM PROCEDURE
1. Press and hold until colorimeter turns on.
2. Press
ENTER
to select TESTING MENU.
3. Select ALL TESTS (or another sequence containing 021 Chromium) from
TESTING MENU.
4. Scroll to and select 021 Chromium from menu.
5. Rinse a clean tube (0290) with sample water. Fill to 10 mL line with sample
water.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use 0.1 g spoon (0699) to add one level
measure of *Chromium Reagent Powder (V-6276). Cap and shake for one
minute. Wait 3 minutes.
8. During the waiting period, fold a piece of lter paper in half, then in half again to
form a cone. Push corners together to open end, and insert into funnel (0459).
9. At the end of 3 minute waiting period, lter sample into a clean tube (0290).
Cap and mix. Insert tube into chamber, close lid and select SCAN SAMPLE.
Record result.
10. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
TOTAL CHROMIUM WITH ACID DIGESTION
PROCEDURE
1. Fill graduated cylinder (0418) to 50 mL line with sample water. Transfer to
Erlenmeyer ask (0431).
2. Use the 1 mL pipet (0354) to add 5 mL ( ve measures) of *Sulfuric Acid, 5N
(7681). Swirl to mix.
NOTE: Highly buffered waters may require pH adjustment. Adjust the pH of
highly buffered samples to 7.0 ±0.5. Continue procedure.
3. Place ask on burner or hot plate. Bring solution to a gentle boil.
4. Fill pipet (0341) with Potassium Permanganate, 0.5% (7682). While gently
swirling ask, add Potassium Permanganate, 0.5% (7682), 2 drops at a time
to boiling solution, until solution turns a dark pink color which persists for 10
minutes. Continue boiling.
5. Add one drop of *Sodium Azide, 5% (7683) to boiling solution. Boil for
approximately 30 seconds. If pink color does not fade, add another drop of
*Sodium Azide, 5%. Continue adding *Sodium Azide, 5% one drop at a time
until pink color disappears.