TEST SEQUENCES
Sequence 1, Sequence 2, And Sequence 3 are alterable sequences. They may
be edited using the Editing Menu. Any of the LaMotte pre-programmed tests or
User Tests may be placed in these sequences in whatever testing order that is
preferred. Some examples of typical sequences are given below.
Sequence 1 Sequence 2 Sequence 3
015 Chlorine F UDV 002 Aluminum 003 Ammonia-N LRF
079 Phosphate HR 035 Cyanuric Acid 032 Cu UDV
009 Benzotriazole 053 Iron Phenanthro 064 Nitrate-N LR
076 pH UDV 055 Manganese LR 067 Nitrite-N LR
12:00:00 001/500 12:00:00 001/500 12:00:00 001/500
061 Molybdenum HR 064 Nitrate-N LR 074 pH PR
086 Silica HR 067 Nitrite-N LR 078 Phosphate LR
045 Hydrazine 077 Phenol 085 Silica LR
032 Cu UDV 078 Phosphate LR
051 Iron Bipyridyl 090 Sul de LR
These alterable sequences allow a series of tests to be setup that are run
frequently. The order of the individual tests in the sequence is determined by the
user. After running a test, press
ENTER
to select the next test in the sequence.
Continue this pattern until the entire sequence has been completed.
All Tests is a xed sequence containing the LaMotte pre-programmed tests, User
Tests, and Absorbance tests.
Modi cation of the alterable sequences is accomplished through the Editing
Menu. This menu is explained in greater detail in Editing Menu (p. 35).
Pressing
EXIT
while in a sequence menu will escape back to the Testing Menu.
Pressing the at any time will turn the colorimeter off.
GENERAL TESTING PROCEDURES
The following are some step by step examples of how to run tests from the
Testing Menu. These test procedures are designed to be used with LaMotte
SMART Reagent Systems.
LaMotte Company continuously updates the list of pre-programmed tests as
the calibrations become available. Pre-programmed calibrations can be added
to the SMART3 Colorimeter in the eld. A Windows-based computer running a
Windows Operating System is required.
Call LaMotte Technical Services at 1-800-344-3100 (410-778-3100 outside the
USA) or email at [email protected] for a current list of available calibrations
and downloading instructions.
26 SMART3 Colorimeter 11.10