PROCEDURE
1. Press and hold until colorimeter turns on.
2. Press
ENTER
to select TESTING MENU.
3. Select ALL TESTS (or another sequence containing 061 Molybdenum HR)
from TESTING MENU.
4. Scroll to and select 061 Molybdenum HR from menu.
5. Fill clean tube (0290) to 10 mL line with sample water.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use a 1.0 mL pipet (0354) to add 1.0 mL of
*Mo Buffer (3997). Cap and mix.
8. Use a second 1.0 mL pipet (0372) to add 1.0 mL of *Molybdenum Oxidizing
Reagent (6485). Cap and mix.
9. Use 0.05 g spoon (0696) to add one measure of Molybdenum Indicator
Powder (6486). Cap and mix until powder dissolves. Solution will turn yellow if
molybdenum is present. Mix the tbe to remove bubbles.
10. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result.
11. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
MOLYBDENUM, High Range SMART3 Test Procedures 11.10