8. Press
ENTER
to save the
character. Repeat the
procedure until the low
concentration value is
complete. After the fi nal
character is complete the
meter will save the low
concentration value. The
screen will display High
Standard for about 1 minute
and the meter will display
the High Standard screen. A
reverse font (dark background
with a light character)
will appear to indicate
the character that will be
adjusted. The same character
will appear in the center of the
display.
High Standard
9 . – 0 1 2 3
0
.0000
12:00:00 001/500
9. Use or and
ENTER
to select the characters for the
high concentration value. In
this example, 7.5 ppm.
High Standard
9 . – 0 1 2 3
7.50000 0
12:00:00 001/500
10. After the fi nal character is
entered the meter will save
the high concentration value.
The screen will display
instructions for completing the
calibration procedure.
Insert Blank
<Enter> continue
12:00:00 001/500
50 SMART3 Colorimeter 11.10
Editing/Set Up