Lamotte SMART3 Colorimeter User Manual Page 32

  • Download
  • Add to my manuals
  • Print
  • Page
    / 294
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 31
4. Press or to scroll
to the desired test factor.
All Tests
001 Alkalinity UDV
002 Aluminum
003 Ammonia-N LRF
004 Ammonia-N LRS
12:00:00 001/500
5. Press
ENTER
to select the test.
002 Aluminum
Scan Blank
Scan Sample
12:00:00 001/500
6. Follow the test procedure
in the manual to test the
prepared standard. Insert the
blank into the chamber. Close
the lid. Press
ENTER
to scan
the blank. The screen will
display Blank Done for about
1 second and then return to
the Test Menu.
002 Aluminum
Scan Blank
Scan Sample
12:00:00 001/500
7. Insert the reacted standard
solution into the chamber.
Close the lid. Press
ENTER
to
scan the sample. The screen
will display Reading for about
1 second. The result will
appear on the screen.
002 Aluminum
0.28 ppm
Scan Blank
Scan Sample
12:00:00 001/500
30 SMART3 Colorimeter 11.10