Lamotte SMART3 Colorimeter User Manual Page 137

  • Download
  • Add to my manuals
  • Print
  • Page
    / 294
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 136
COBALT
PAN METHOD · CODE 4851
QUANTITY CONTENTS CODE
60 mL *Cobalt Buffer *4852-H
60 mL *Cobalt Indicator Reagent *4853-H
30 mL *Stabilizer Solution *4854-G
2 Pipet, 1.0 mL, plastic 0354
1 Pipet, 0.5 mL, plastic 0353
*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents
go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail,
phone or fax.
Cobalt rarely occurs in natural water. It is used in the manufacture of alloys
to increase corrosion resistance and strength. It is found in wastewaters as a
corrosion by-product.
APPLICATION: Industrial wastewater.
RANGE: 0.00–2.00 ppm Cobalt
MDL: 0.04 ppm
METHOD: PAN (1-[2-Pyridylazo]-2-Naphthol) forms a greenish
complex with Cobalt (Co
+2
) at a pH of 5.
SAMPLE HANDLING
& PRESERVATION:
Store samples in acid-washed plastic bottles. Adjust
pH to less than 2 with nitric acid. Adjust sample pH to 5
before testing.
INTERFERENCES: Iron (+2) and high concentrations of heavy metals.
SMART3 Test Procedures 11.10 COBALT
Test Procedures
Page view 136
1 2 ... 132 133 134 135 136 137 138 139 140 141 142 ... 293 294