Lamotte SMART3 Colorimeter User Manual Page 168

  • Download
  • Add to my manuals
  • Print
  • Page
    / 294
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 167
PROCEDURE
1. Press and hold until colorimeter turns on.
2. Press
ENTER
to select TESTING MENU.
3. Select ALL TESTS (or another sequence containing 035 Cyanuric Acid) from
TESTING MENU.
4. Scroll to and select 035 Cyanuric Acid from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter and pour out water. Use a graduated cylinder or
similar to measure 5 mL of sample water and pour into colorimeter tube.
8. Use the 5 mL syringe (0807) to add 5 mL of *Cyanuric Acid Test Solution
(4856). Cap and mix thoroughly. A precipitate will form if cyanuric acid is
present. Wait 1 minute.
NOTE: This reagent bottle has a special tting which enables the syringe to
be inserted into the top of the bottle. With syringe in place, invert bottle and
withdraw syringe plunger until 5 mL of reagent is contained in the syringe
barrel. Remove syringe from reagent bottle and depress plunger to dispense
into the tube.
9. At end of 1 minute waiting period, mix thoroughly, insert tube into chamber,
close lid and select SCAN SAMPLE. Record result.
10. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
NOTE: For the most accurate results, the sample and reagents should be at 25
±4°C.
CYANURIC ACID, Liquid SMART3 Test Procedures 11.10
Test Procedures
Page view 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 293 294

Comments to this Manuals

No comments