APPLICATION: swimming pools, surface and saline waters, hydroponic
solutions, industrial waste.
RANGE: 0.00 – 80.0 ppm boron
MDL: 5 ppm
METHOD: Borate reacts with a selective boron indicator powder to
form a colored complex at pH 7.2 to 7.4 in proportion to
the concentration of boron present.
SAMPLE HANDLING &
PRESERVATION:
Store samples in polyethylene bottles. Do not use
borate detergents or glassware.
INTERFERENCES: Interferences in swimming pool water are unlikely.
BORATE, UDV SMART3 Test Procedures 11.10