  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 55
8. Press
ENTER
to save the
character. Repeat the
procedure until the K0 value
is complete. After the nal
character is complete the
meter will save the K0 value
and the meter will display K1
screen.
Enter K1
9 . – 0 1 2 3
0
.00000
12:00:00 001/500
9. Use , and
ENTER
to select the characters for
the remaining values: K1, K2,
K3, and over range. In this
example, 10 ppm.
Enter Overrange
9 . – 0 1 2 3
10.00000 0
12:00:00 001/500
10. After the nal character is
entered the meter will save
the constants. The screen will
display Storing... and return
to the Edit Test menu.
111 H2O
Name the Test
Select Vial/WL
STD Calibration
Enter Constants
12:00:00 001/500
54 SMART3 Colorimeter 11.10