SELECTING THE NUMERICAL FORMAT OF THE
RESULT
To input tests with very different ranges, the number of decimal places displayed
for a result can be selected. A test which ranges from 20 to 1000 ppm should
not be displayed with three decimal places. A test with a range from 0.010 to
0.500 needs three decimal places (the microprocessor will always calculate
the concentration to many more signi cant gures than will be displayed). The
choice of 0, 1, 2, or 3 decimal places are available.
1. From the User Test menu,
press or to scroll
to Decimal Places.
111 H2O
Select Vial/WL
STD Calibration
Enter Constants
Decimal Places
12:00:00 001/500
2. Press
ENTER
to select Decimal
Places.
Decimal Places?
None 0
One 0.0
Two 0.00
Three 0.000
12:00:00 001/500
3. Press or to scroll
to the desired number of
decimal places.
Decimal Places?
None 0
One 0.0
Two 0.00
Three 0.000
12:00:00 001/500
4. Press
ENTER
to select the
decimal places. The screen
wil display Storing... for about
1 second and the meter will
return to the Edit Test menu.
111 H2O
Select Vial/WL
STD Calibration
Enter Constants
Decimal Places
12:00:00 001/500